HydroSCOUT ANALYZER System PPM RANGE for LUBRICATING OILS (HS-MTR-02)

Analytes |
Water |
Matrix |
Lubricating Oils, Industrial Oils, BioDiesel |
Detection Method |
Quantitative calcium Hydride reaction |
Action Levels |
50 - 10,000 ug/mL |
MDL |
50 ug/mL |
MQL |
150 ug/mL |
Interferences |
Ethylene Glycol / Acids |
Overall Accuracy |
10% +/- MDL |
Analysis Time |
Less than ten minutes |
Catalog Number |
HS-MTR-02 |
Packaging |
Ships as single unit |
Analyzer system for quantifying moisture content of industrial
oils - ppm range.
On-Site Test for Water in Industrial/Lubricating Oils &
BioDiesel
U.S. EPA SW-846 Draft Method 9001
Moisture in oil is an important parameter that must be addressed
when considering contaminates in lubricating and hydraulic oils.
Analyses performed in the laboratory are time consuming and costly.
Turnaround time for most water analyses is 1 to 2 weeks.
HydroSCOUT is an on-site test that quantifies water in industrial
oils quickly and easily. The HydroSCOUT method incorporates the
standard reaction of water with calcium hydride to produce one mole
of hydrogen for every mole of water. By inserting the reaction tube
into the hand-held meter, the instrument converts the pressure of
the hydrogen into the amount of water in the sample. The results
are reported in ug/mL. To convert to parts per million, simply
divide your results by the specific gravity of the oil
sample.
HydroSCOUT METER
PPM Range
The meter is menu driven for ease of use. All programs are
auto-calibrated and perform quality control checks to minimize
false negatives and ensure accuracy.
PPM Range Reagents
All reagents are premeasured and sealed in glass ampules for safe,
accurate and consistent results. Analysis time is less than 10
minutes utilizing a sampling syringe, one test tube and one
breakable ampule. Each test comes with a disposal ampule that
allows easy disposal of used tests in normal waste. This
environmentally safe test kit comes packaged with 40 tests per
box.
Technical Hint
Videos
HydroSCOUT PPM - Assembling The Reaction Tube
Related Papers (PDF Downloads)
Manuals:
- Hydroscout User's Manual - July
13, 2006 || Download
MSDS:
- Hydroscout Oil/Soil - December
11, 2010 ||
Download
FAQs:
- Hydroscout System Questions and
Answers - June 1, 2006 || Download
Technical Papers:
- U.S. EPA SW-846 Method 9001
(DTP-17-01) - July 13,
2006 || Download
|